SERVICE
GUIDE
SUPPORT: TOLL-FREE 888-383-DIVE │ FAX 510-729-5115 │ E-MAIL [email protected]
©
2002 DESIGN
(
2012)
8. Remove the
ELECTRICAL CONNECTORS (5)
from the
OXYGEN SENSORS (7)
.
NOTE:
A small flat blade screwdriver, preferably a jeweler's screwdriver, may be used to release the
electrical connector locking mechanism without damaging the connectors.
9. Inspect the
SENSOR (7)
contacts for damage or corrosion. Clean or discard the sensor as needed.
CAUTION: The electrical connectors have thin precious metal plating. DO NOT scrape corrosion from
contacts with a hard sharp object. DO NOT use abrasives, petroleum distillates, silicone based clean-
ers, erasers, combustible chemicals, or any other harsh cleaners. Hollis recommends the use of DeoxIT
Gold® GN5 for cleaning the OXYGEN SENSOR (7) electrical contacts and the contacts on the WIRE
HARNESS (5).
10. Being careful not to physically damage the electrical connectors, remove and inspect the
WIRE HARNESS
(5)
for corrosion and damage. Clean or discard as appropriate.
11. Inspect the
BULKHEAD CONNECTOR(4)
, that the
WIRE HARNESS (5)
plugs into, on the
SCRUBBER
HEAD (1)
for corrosion or damage. Clean as appropriate.
MILLIVOLT/LINEARITY TEST & CLEANING
WARNING: Any SENSORS (7) being used in the PRISM 2 unit must pass this Millivolt/Linearity Test.
TECHNICIANS ONLY:
NOTE:
Any opened SENSORS (7) must have serial numbers, date of testing, and mV reading data re-
sults documented for the customer on a copy of the Hollis Rebreather Receiving and Final Inspection
Form. You should retain the original copy for your records.
WARNING: The SENSORS (7) must be installed in the PRISM 2, tested, and pass the Final Inspection
Procedure before returning the PRISM 2 unit to your customer for use. The Final Inspection Proce-
dure may be found in the Receiving and Final Inspection Section of the PRISM 2 Service Guide.
1. Visually inspect the
OXYGEN SENSOR'S (7)
electrical terminal, Teflon membrane, and plastic body for dirt,
corrosion, or physical damage.
2. The terminals may be cleaned using Deoxit Gold
™
GN5. Use the chemical sparingly and wipe any excess
away with a clean lint free cloth. If the terminals are physically damaged the sensor
MUST NOT BE USED
.
3. The plastic body of the
SENSOR (7)
should be wiped clean with a lint free cloth. If needed, sparingly apply
Blue Gold
®
to the lint free cloth to clean the plastic
SENSOR (7)
body
ONLY
.
DO NOT
expose the
SENSOR (7)
membrane or electrical terminals to the cleaner.
4. The Teflon membrane may be gently rinsed with distilled water
ONLY
if salt crystals are present. It must then
be allowed to air dry.
WARNING: DO NOT use compressed air to dry the delicate membrane surface, or the SENSOR (7)
may be damaged.