5 - 38
Code Error
Message
Countermeasure
1231 Time has run out.
“Auto Start” cannot be started.
This message appears if a time-out (40 s) has
occurred during HV auto start (HV interlink:
HV-ON/AFC/ABC). Then, auto start will stop.
Click the [OK] button to close the message.
Restart the system by the following procedure.
1.
Turn off the PC power.
2.
Turn off the main unit power.
3.
Restart the PC.
4.
Turn on the main unit power and start the
application.
If the same message appears repeatedly after
restart of the system, contact the nearest service
representative.
1241 Time has run out.
“Condition Setting” cannot be
executed.
This message appears if a time-out (60 s) has
occurred during setting of beam axis alignment
conditions. Then, beam axis alignment stops.
Click the [OK] button to close the message.
Restart the system by the following procedure.
1.
Turn off the PC power.
2.
Turn off the main unit power.
3.
Restart the PC.
4.
Turn on the main unit power and start the
application.
If the same message appears repeatedly after
restart of the system, contact the nearest service
representative.
1242
Beam Axis Alignment uses the
BSE.
Was the reference sample set
in the center of the stage?
This message appears to urge the user to
confirm that the reference sample is set to the
center of the specimen stage before beam axis
alignment starts. If the specimen is already set,
click the [OK] button. Beam axis alignment will
be started using the BSE. If the specimen is not
yet set, click the [Cancel] button.
1243 After setting the reference
sample in the center of the
stage, make Beam Axis
Alignment again.
This message appears when the “Cancel”
button is clicked in response to the message
1242. After setting the specified reference
sample on the specimen stage, retry beam axis
alignment.