5 - 42
Code Error
Message
Countermeasure
1530 Evacuation System is wrong.
Start up the main unit and
application again.
Contact service if the failure
occurs repeatedly.
This message appears when the evacuation
SEQ did not run successfully due to an error
other than the above (1510 ~ 1514) in the
evacuation system.
If the same message appears repeatedly after
restart of the system, contact the nearest service
representative.
1570 The power supply of -15V for VA
was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F5 on the CONTROL0 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1571 The power supply of +15V for
VA was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F4 on the CONTROL0 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1572 The power supply of -15V for
SG was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F3 on the CONTROL0 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1573 The power supply of +15V for
SG was shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (F2 on the CONTROL0 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
1574 The power supply of +24V was
shut down.
Start up the main unit and
application again.
An overcurrent in the power supply or a blown
fuse (on the IO-DRV00 board, F5, F6, F7, F8, or
F9, or more than one fuse) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.