4 - 28
(3) Assembly of Flow Cell
(a) Assemble the flow cell into the cell holder in the
orientation shown in Fig. 4-13 and fix it with the cell
clamp.
(b) Tighten the upper and lower retaining screws 1 and 2.
Fig. 4-13 Assembly of Flow Cell Unit
(c) Attach the cell window and shielding plate.
(d) After assembling the flow cell in the order reverse to
disassembly, flow a liquid through the unit and make
sure there is no leakage.
NOTE
: Be careful not to scratch the flow cell surface when
assembling it.
NOTE
: Take care about the orientation of the polished
surfaces of the flow cell when putting it into the cell
holder (see Fig. 4-13). A mistake in the
orientation may preclude obtaining the proper
performance from the instrument.
NOTE
: Tighten setscrew 1 fully with hand, and then
tighten setscrew 2 about 30
manually or tighten
setscrew 1 further 30
with a wrench.
Excitation light
Enlarged view of flow cell
Transmitted
excitation light
(Instrument front)
Polished surfaces
(with slit)
Emission
View from top
Hg light
entrance
Flow cell
Reflection
plate
(Instrument rear)
(Instrument front)
(Instrument rear)
Polished surface
(without slit)