(5) Setting of second and subsequent steps of time program:
(a) Set the TIME to be changed
.
<Step setting screen (TIME indicated)>
TIME
Ex
Em
BASE
PMT
_ 10.0
1
1
Press numeric keys 0 to 6 0 0 and then the ENT key.
* Setting range: 0 to 600 minutes, in increments of 0.1
minute
The cursor now moves to Ex.
(b) Set an Ex (excitation) wavelength to be changed.
<Step setting screen (Ex indicated)>
TIME
Ex
Em
BASE
PMT
10.0
400
1
1
Press numeric keys 2 0 0 to 8 5 0 and then the ENT
key.
The cursor next moves to Em.
(c) Set an Em (emission) wavelength to be changed.
<Step setting screen (Em indicated)>
TIME
Ex
Em
BASE
PMT
10.0
400
500
1
Press numeric keys 2 5 0 to 9 0 0 and then the ENT
key.
The cursor next moves to BASE.