3 - 35
Press
PROG
key.
SET PROGE NO. (1 - 9)
1
Press 1 key and then the ENT key.
EDIT MODE (MODIFY=0, NEW=1)
0
Press 1 key and then the ENT key.
TIME
Ex
Em
BASE
PMT
0.0
1
1
Press 0 key and then the ENT key.
Press 2 5 0 keys and then the ENT key.
Press 3 5 0 keys and then the ENT key.
Press 1 key and then the ENT key.
Press 3 key and then the ENT key.
TIME
Ex
Em
BASE
PMT
0.0
250
350
1
3
1
Press 1 0 keys and then the ENT key.
Press 2 7 0 keys and then the ENT key.
Press 3 7 0 keys and then the ENT key.
Press 2 key and then the ENT key.
Press 2 key and then the ENT key.
TIME
Ex
Em
BASE
PMT
10.0
270
370
2
2
1
Press 2 0 keys and then the ENT key.
TIME
Ex
Em
BASE
PMT
20.0
1
Press the ESC
key.
Time program setting is now finished.
TIME
Ex
Em
FL
PROG
LAMP
[I]
0.0
250
350
0.0
1
Xe
Press the START key.
The time program starts and measurement is initiated.