4 - 17
Wavelength calibration is conducted by entering a
new offset value (value to which the deviation is
added).
Press numeric keys (from 1 to 2 0) and then the
ENT key.
*
Input range for offset value: 1 to 20
* Calculation of new offset value:
New offset value = present offset value
–
wavelength deviation
Example: 12 = 10
– (-2)
(iii) The 1
st
monitor screen appears.
(b) Calibration of emission wavelength
(i)
Press the UTLTY
key, and the utility setting
screen appears.
<Utility setting screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
98
USE PROG=3, PMT VOLT=4, OTHERS=5)
Press numeric keys 9
8 and then the ENT
key.
(ii) The emission wavelength calibration mode
screen appears.
<Emission wavelength calibration mode screen>
Em WL ADJUST (1 - 20)
10 (ONLY FOR FACTORY USE) (-2)
The present offset value is indicated at Em WL
ADJUST. Indicated at the bottom right of the
screen is the wavelength check result (deviation
of wavelength; indication range of
–10 to 10).
Wavelength calibration is conducted by entering a
new offset value (value to which the deviation is
added).
Press numeric keys (from 1 to 2 0) and then the
ENT key.
* Input range for offset value: 1 to 20
* Calculation of new offset value:
New offset value = present offset value
–
wavelength deviation
Example: 12 = 10
– (-2)