146
Measuring Object Short Circuit due to Relay Contact Welding
Module fuse rating
The SW9001 and SW9002 are provided with a fuse (board mounted type) in HIGH and LOW for
each channel input to prevent possible risks if a high current flows through the circuit due to relay
contact welding.
Module
Channel
Rated current
Fusing current
SW9001
All channels
1.6 A DC
4.0 A DC
SW9002
Sense channel
1.6 A DC
4.0 A DC
Source channel
2.5 A DC
6.25 A DC
Return channel
2.5 A DC
6.25 A DC