121
SW1001, SW1002 Functional Specifications
9.3
SW1001, SW1002 Functional Specifications
Channel switching
The specified slot and channel is connected to the analog bus based on the
communication command.
Connection method
2-wire
Connect to
TERMINAL 1
4-wire
Connect to
TERMINAL 2
4-terminal pair
Connect to
TERMINAL 3
Connect to
TERMINAL 2
and
TERMINAL 3
in parallel inside.
Default setting
SW9001: 2-wire
SW9002: 4-terminal pair
Channel delay
The delay time allowed to elapse from channel relay close operation to outputting of the
EXT. I/O CLOSE signal can be set.
Delay time
0.000 s to 9.999 s
Default setting
Delay time: 0.000 s
Shield switching
The shield wire connection point can be set for each slot.
However, it is reconfigured to the default with each cycle, according to the connection
method when the connection method is changed.
Connection destination
Default according to the connection method
Not connected
–
TERMINAL 1
LOW terminal
2-wire
TERMINAL 2
source LOW terminal
(only for model SW9001)
–
TERMINAL 3
sense shield terminal
4-terminal pair
GND
4-wire
TERMINAL 1
LOW te
TERMINAL 3
sense shield terminal (only for model SW9001)
–
Scan function
Operation
Switch the channel according to the created scan list.
Scan operation is triggered by the EXT. I/O SCAN signal or
*
TRG
command input.
Scan operation
Trigger source
STEP
Switches channels in the order specified by the scan
list at each signal input.
EXT. I/O can be used to synchronize channel
switching and trigger measurement with one
measuring instrument.
Scan list
Registers scan channels.
Maximum number of steps: 1000
Default setting
Scan operation trigger source: STEP
Scan list: none
Communication
command forward
Communication data can be forwarded (sent/received) between the host (PC, PLC) and
the instrument connected to the
INSTRUMENT RS-232C
connector of the device.
(Communications with the instrument can be performed using the communication port.)
Setting backup
Backs up communication settings and scan lists in the non-volatile memory.
9
Specifications
Summary of Contents for SW1001
Page 2: ......
Page 42: ...38 Precautions for Measurement...
Page 48: ...44 Resetting Scan Operation...
Page 50: ...46 Checking the Device Status...
Page 52: ...48 Initialization Settings...
Page 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Page 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Page 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Page 158: ...154 Outline Drawings...
Page 161: ...16 01 EN...
Page 162: ......
Page 163: ......
Page 164: ......