97
Message Reference
(3) Shield switching
Set and query the shield switching
Syntax
Command
:SYSTem:MODule:SHIeld <
Slot No.
>,<
Shield wire connection
destination
>
Query
:SYSTem:MODule:SHIeld?
Slot No.
>
Response
<Shield wire connection destination>
<Slot No.> = 1 - 3 (NR1) (SW1001)
1 - 12 (NR1) (SW1002)
<Shield wire connection destination> = OFF/GND/TERMinal1/
TERMinal2/TERMinal3/T1T3/SNS2L
OFF: Not connected
GND: Connected to ground
TERMinal1:
TERMINAL 1
, connect to LOW terminal
TERMinal2:
TERMINAL 2
,
TERMINAL 3
, connect to source LOW
terminal
TERMinal3:
TERMINAL 3
, connect to sense shield terminal
T1T3:
TERMINAL 1
, LOW te
TERMINAL 3
, connect to sense
shield terminal
Description
Sets the shield wire connection destination for each slot.
The settings vary depending on the module type.
When the shield switching is set, all relay is opened.
Module
Parameter that can be set
Connection destination terminal
SW9001
Off
Not connected
GND
GND
TERMinal1
TERMINAL 1
, LOW terminal
TERMinal2
TERMINAL 2
,
TERMINAL 3
, source LOW
terminal
TERMinal3
TERMINAL 3
, sense shield terminal
T1T3
TERMINAL 1, LOW te TERMINAL 3,
sense shield terminal
SW9002
Off
Not connected
GND
GND
TERMinal1
TERMINAL 1
, LOW terminal
TERMinal3
TERMINAL 3
, sense shield terminal
Example
:SYST:MOD:SHI 1,GND
Connect the shield wire for
SLOT 1
to GND.
:SYST:MOD:WIRE:MODE? 5
OFF
The shield wire for
SLOT 5
is not connected.
Note
• When the connection method is selected, the shield switching is set to the specified
value. Set the shield switching again as necessary.
•
The [
TERMINAL 1
, LOW te
TERMINAL 3
, sense shield terminal (T1T3)]
shielding connection on the SW9001 should only be used when connecting a BT3562
and DM7275/DM7276 at the same time while the DM7275/DM7276 contact check
function is not operating properly.
• Specifying a slot without a module will result in an execution error.
• An execution error occurs during scan operation (command).
8
Communication Function
Summary of Contents for SW1001
Page 2: ......
Page 42: ...38 Precautions for Measurement...
Page 48: ...44 Resetting Scan Operation...
Page 50: ...46 Checking the Device Status...
Page 52: ...48 Initialization Settings...
Page 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Page 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Page 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Page 158: ...154 Outline Drawings...
Page 161: ...16 01 EN...
Page 162: ......
Page 163: ......
Page 164: ......