106
Message Reference
Query relay opening/closing frequency
Syntax
Query
:SYSTem:MODule:COUNt? <
Slot No.
>
[,<Relay No.>]
Response
<Slot No.> = 1 to 3 (NR1) (SW1001)
1 to 12 (NR1) (SW1002)
<Opening/closing frequency> = 0 to 1000000000 (NR1)
Description
Returns to the specified relay opening/closing frequency.
Example
:SYSTem:MODule:COUNt? 1
101257
Note
• The relay opening/closing frequency is saved in the non-volatile memory inside each
module. If no relay number is specified, the query will return the count for the relay for
the selected slot that has been turned off and on the most times. (See: p. 134, p. 140)
• Specifying a slot without a module will result in an execution error.
(11) Initialization
Initialize the device
Syntax
Command
:SYSTem:PRESet
:STATus:PRESet
Description
Resets the device to its initial state.
See: "6.1 Initialization Settings" (p. 47)
The operation is the same for
:SYSTem:PRESet
and
:STATus:PRESet
.
Example
:SYST:PRES
Note
• The communication conditions are not initialized.
•
All closed channels are opened.
• Initialized settings are not preserved when the power is turned off. To save initialized
settings, execute the
SYSTem:BACKup
command.
See: "Backing up settings" (p. 105)
(12) Test
Relay contact welding test
Syntax
Command
:TEST:RELayshort <
Slot No.
>,<
Test item
>
<Slot No.> = 1 to 3 (NR1) (SW1001)
1 to 12 (NR1) (SW1002)
<Test item> = CH/BUS/OPEN
CH:
Channel relay test
BUS:
Bus relay test
OPEN:
Open the relay after all test is finished.
Description
Execute relay contact welding test to each slot.
The test must be executed to both channel relay and bus relay.
For details, see "Checking for relay contact welding" (p. 26)
Example
:TEST:RELayshort 1,CH
Execute the test to channel relay of
SLOT 1
.
Summary of Contents for SW1001
Page 2: ......
Page 42: ...38 Precautions for Measurement...
Page 48: ...44 Resetting Scan Operation...
Page 50: ...46 Checking the Device Status...
Page 52: ...48 Initialization Settings...
Page 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Page 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Page 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Page 158: ...154 Outline Drawings...
Page 161: ...16 01 EN...
Page 162: ......
Page 163: ......
Page 164: ......