Command Message List
8
1-2. Output monitor
Command
Description
Formats
MON
Set the monitor ON or OFF
d1 (Setting of monitor: 0 to 1)
0: Monitor OFF
1: Monitor ON
[Format]
MON d1
d1: NR1 format
[Response]
OK or CMD ERR
MON?
Query the setting of monitor
The contents of responses are the same as the settings.
[Format]
MON?
[Response]
d1: NR1 format
RMV?
Query the current monitor value during output of voltage
This command is valid only in CV:2.5V or CV:25V.
d1 (Current monitor value: unit; mA)
-28.00 to 28.00
[Format]
RMV?
[Response]
d1: NR2 format or
CMD ERR: When the input is
out of the measure-
ment range.
RMC?
Query the voltage monitor value during output of current
This command is valid only in CC:25mA.
d1 (Voltage monitor value: unit; V)
-28.00 to 28.00
[Format]
RMC?
[Response]
d1: NR2 format or
CMD ERR: When the input is
out of the measure-
ment range.
RMT?
Query the temperature monitor value
This command is valid only in TC:0°C or TC:RJ.
d1 (Temperature monitor value (in TC:RJ): unit; °C)
-25.0 to 80.0
d1 (Temperature monitor value (in TC:0°C): unit; °C)
0.0
[Format]
RMT?
[Response]
d1: NR2 format or
CMD ERR: When the input is
out of the measure-
ment range.