Examples of sending commands
12
1. Sourcing
1-1. Normal sourcing
1-2. Output monitor
1-3. Memory sourcing
2. Measurement
Examples of sending commands
When sourcing 4 mA
Command
Description
FCC 2<CR><LF>
Sets [
CC:25mA
].
CCA 4<CR><LF>
Sets the setting value to 4 mA.
OUT 1<CR><LF>
Sets output to ON.
OUT 0<CR><LF>
Sets output to OFF.
When monitoring the output current while sourcing 24 V.
Command
Description
FCC 1<CR><LF>
Sets [
CV:25V
].
CVV 24<CR><LF>
Sets the setting value to 24 V.
OUT 1<CR><LF>
Sets output to ON.
MON 1<CR><LF>
Switches monitor on.
RMV?<CR><LF>
Query the output current.
MON 0<CR><LF>
Switches monitor off.
OUT 0<CR><LF>
Sets output to OFF.
When sourcing the setting value in address 01 with recall sourcing in the current
source function.
Command
Description
FCC 2<CR><LF>
Sets [
CC:25mA
].
MMD 1<CR><LF>
Sets recall source mode.
RCL 1<CR><LF>
Sets the address to 1.
OUT 1<CR><LF>
Sets output to ON.
OUT 0<CR><LF>
Sets output to OFF.
When performing voltage measurement at [
V:25V
]
Command
Description
FCM 2<CR><LF>
Sets [
V:25V
].
ADJ<CR><LF>
Execute the zero adjustment. Short out the ends of the test
leads.
RDV?<CR><LF>
Query the measured voltage value.