48
Basic Specifications
Functions Specification
Voltage output function
Operating method
Sink/source (support for charging and discharging)
Output voltage setting
value
1.0 V to 1000.0 V (set in 0.1 V increments; valid setting range varies by model)
Refer to “Configuration” (p. 46) in “5.2 Basic Specifications”)
Generation control
Output only when EXT I/O interface’s
OUTPUT signal is on.
Output on/off
Can be set independently for each channel.
Voltage error alarm
Operation: Alarm is generated when the monitor voltage falls outside the set
range.
Valid setting range : ±2% to ±19% (1% resolution)
Current limit function
Limit method
Current can be limited independently for each channel.
Limit value setting
2 mA to 50 mA (set in 1 mA increments; valid setting range varies by model)
Refer to “Configuration” (p. 46) in “5.2 Basic Specifications”)
Current limit direction
Current can be limited in both directions.
Backup Function
Backup items
Output voltage setting, alarm setting, GP-IB address
Voltage monitor function
Voltage monitor
Measures and displays output voltage for each circuit.
Interlock function
Operation
Interlock is enabled when set to “on,” with low input, or when terminals are
shorted.
Measurement is disabled, including by device keys and communications
commands, when interlock is enabled.
Output setting function for the device interlocked
Operation
The output setting with the device interlocked can be chosen between an
impedance state and a discharging state.
Setting procedure
Cannot be set by sending commands.
Press and hold both the scroll keys (
[PAGE UP▲]
and
[PAGE DOWN▼]
)
on the right of the screen simultaneously for 3 s.
The GP-IB-address assigning screen is displayed.
Press and hold both the scroll keys (
[PAGE UP▲]
and
[PAGE DOWN▼]
)
again simultaneously for 3 s.
The setting screen is displayed.
Press either of the scroll keys (
[PAGE UP▲]
or
[PAGE DOWN▼]
) to set the
GP-IB-address and turn off the device to confirm the address you entered.
Supported model
Hioki Model SM7420 Super Megohm Meter,