34
Message List
3.6 Message List
RS-232C-only commands are indicated by
RS-232C
. When using the RS-232C interface to send
commands, include a uniform wait time of 100 ms (excluding the following exceptions).
Command
Description
Format
Communication condition
RMT
Remote switching request
RS-232C
[Format] RMT
Delimiter
DLM
Talker delimiter specification
d1 (delimiter specification: 0 to 2)
0: LF<EOI> Default
1: CRLF<EOI>
2: <EOI>
Note: This setting reverts to its default when the device is powered on.
A combination CR+LF is used as the RS-232C delimiter for both
data transmission and reception.
[Format] DLM d1
d1: NR1
format
DLM?
Delimiter query
The contents of responses are the same as the settings.
[Format]
[Response]
DLM?
d1
Output voltage setting
VAI
Circuit A output voltage setting
d1: 1.0 V to 1000.0 V
Note: Set as an absolute value.
The valid setting range varies with the model.
[Format] VAI d1
d1: NR2
format
VAI?
Circuit A output voltage setting query
The contents of responses are the same as the settings.
[Format]
[Response]
VAI?
d1
VBI
Circuit B output voltage setting
d1: 1.0 V to 1000.0 V
Note: Set as an absolute value.
The valid setting range varies with the model.
[Format] VBI d1
d1: NR2
format
VBI?
Circuit B output voltage setting query
The contents of responses are the same as the settings.
[Format]
[Response]
VBI?
d1
Output voltage setting range by model
Model
Lower limit (V) Upper limit (V)
SM7860-51, -61
1.0
500.0
SM7860-52, -62
250.0
1000.0
SM7860-53, -63
1.0
500.0
SM7860-54, -64
250.0
1000.0
SM7860-55, -65
1.0
500.0
SM7860-56, -66
250.0
1000.0
SM7860-57, -67
1.0
10.0
SM7860-58, -68
1.0
500.0
Alarm setting
ARM
Voltage error alarm setting
d1: 2% to 19%, Circuit A voltage error alarm setting
d2: 2% to 19%, Circuit B voltage error alarm setting
Note: d1 or d2 can be omitted.
[Format] ARM d1,d2
ARM d1
ARM ,d2
ARM?
Voltage error alarm setting query
The contents of responses are the same as the settings.
[Format]
[Response]
ARM?
d1,d2
Summary of Contents for SM7860 Series
Page 2: ......
Page 4: ...Contents ii...
Page 12: ...8 Operating Precautions...
Page 26: ...22 Output Setting for the Device Interlocked...
Page 61: ......
Page 62: ......
Page 63: ......
Page 64: ......