17
Connecting the Device to the Measuring Instrument
2.3 Connecting the Device to the Measuring
Instrument
WARNING
To prevent an electric shock or damage to the equipment, always observe the following
precautions when connecting to voltage output terminal:
•
Always turn off the device and any devices to be connected before connecting the
voltage output connectors.
CAUTION
To avoid damage to the device, be sure to observe the following precautions:
•
Do not connect the device to a load with a charge in excess of the output voltage range.
•
Do not connect the device to a load with a charge that has the opposite polarity of the output
voltage range.
•
Do not connect the device to a load that has a charge when the device is turned off.
Connection Method
2
1
1
Confirm that the device is turned off.
2
Connect model L2221 Connector*
1
(optional) to the voltage output
terminal on the rear of the device.
This completes the process of
connecting the device to the
measuring instrument.
*1: Refer to “Voltage output pin assignment specifications” described below to modify the L2221 to
connect measurement targets.
Voltage output pin assignment
specifications
1
2
3
4
5
6
7
9
10
8
Pin No.
OUT1
Circuit A
OUT2*
1
Circuit A
OUT3
Circuit B
OUT4*
2
Circuit B
1
CH1
CH1
CH1
CH1
2
CH2
CH2
CH2
CH2
3
CH3
CH3
CH3
CH3
4
CH4
CH4
CH4
CH4
5
CH5
CH5
CH5
CH5
6
CH6
CH6
CH6
CH6
7
CH7
CH7
CH7
CH7
8
CH8
CH8
CH8
CH8
9
NC
NC
NC
NC
10
COM
COM
COM
COM
*1: On the SM7860-55, -56, -65, and -66, serves as the discharging terminal.
*2: On the SM7860-55, -56, -57, -58, -65, -66, -67, and -68, serves as the
discharging terminal.