21
Output Setting for the Device Interlocked
2.7 Output Setting for the Device Interlocked
This section describes how to choose between an impedance state and a discharging state for the output
with the device interlocked.
This setting remains the same even after the device is initialized.
ILOCK DCHG: OFF
High-impedance state (Default)
ILOCK DCHG: ON
Discharging state
1
Press and hold both of the scroll keys (
[PAGE UP▲]
and
[PAGE DOWN▼]
) on the right of
the screen simultaneously for approx. 3 s.
(The address can be set from the P1 or P2 screen.)
The GP-IB-address assigning screen is displayed.
V A ( + ) I R : 1 0 0 0 . 0
V B ( + ) I R : 2 5 0 . 0
T E R M I N A L :
O U T 1 : 1 1 1 0 0 0 0 0
O U T 2 : 1 1 1 0 0 0 0 0
O U T 3 : 1 1 1 0 0 0 0 0
O U T 4 : 1 1 1 0 0 0 0 0
V M A : 1 0 0 0 . 0 O K
V M B :
2 5 0 . 0 O K
G P I B A D D R : 1
3 s
C L M 1 : 1 0 m A
C L M 2 : 1 0 m A
C L M 3 : 1 0 m A
C L M 4 : 1 0 m A
2
Press and hold both the scroll keys (
[PAGE UP▲]
and
[PAGE DOWN▼]
) again
simultaneously for 3 s.
The setting screen for the output setting for the device interlocked is displayed.
3
Press either of the scroll keys (
[PAGE UP▲]
or
[PAGE DOWN▼]
).
The output state is switched.
I L O C K D C H G : O N
I L O C K D C H G : O F F
4
Turn off the device once you have finished configuring the settings.