71
Setting Measurement Conditions
LR8512 Wireless Pulse Logger
Making settings for integrated measurement
The number of integrated pulses output from the integrated power meter or flow
meter is measured.
1
2
3
5
4
1
:
ON
1
Select the channel to be set and check the Measurement checkbox
[
]
(ON).
2
Select the input type.
Settings:
Instantaneous
Measures the number of pulses input to the instrument within the
recording interval.
The number of pulses is reset for each recording interval.
Add
Measures the number of integrated pulses after measurement starts.
Addition
Instantaneous
HIGH
LOW
Detection
level
3
Select the measurement reference (slope).
Settings:
Rise
Integrates the number of times the pulse changes from LOW to HIGH.
Fall
Integrates the number of times the pulse changes from HIGH to LOW.
4
Select the HIGH/LOW reference value (detection level).
Settings:
1 V
Determines 1.0 V or higher to be HIGH, 0 V to 0.5 V to be LOW.
4 V
Determines 4.0 V or higher to be HIGH, 0 V to 1.5 V to be LOW.