PEM200 User Manual
38
Command
:DETector:Gain #
Short
command
:DET:G #
Type
Get/Set
Purpose
Sets the gain level of the synchronous detector. The gain range is 0.25 to
40000 . The gain circuitry only has 8 selectable gain levels and the closest
option will be chosen and returned.
Input example
:DET:G 20
Output
example
[GAIN](4.2)
Command
:DETector:HARMonics #, #
Short
command
:DET:HARM #, #
Type
Get/Set
Purpose
Select the channel and harmonic of interest to measure when the
synchronous detector is activated. The range of the harmonic is 1 to 5.
Input example
:DET:HARM 1, 2 (Channel 1, 2f)
Output
example
[HARMONIC](1,2)