PEM200 User Manual
37
Command
:MODulator:STABLE?
Short
command
:MOD:STABLE?
Type
Get
Purpose
Returns if the PEM is stable (binary)
Input example
:MOD:STABLE?
Output
example
[STABLE](1)
Command
:DETector:SampleFREQuency #
Short
command
:DET:SFREQ #
Type
Get/Set
Purpose
Sets the sampling frequency of the synchronous detector in Hz. The valid
range is 1MHz (1e6) to 10MHz (1e7). Suggested and default sample
frequency is the suggested 10MHz.
Input example
:DET:SFREQ 1e6
Output
example
[SCLOCK](1.6)