PEM200 User Manual
35
Command explanation
Commands are listed below. The input command is listed in full with the symbol #
denoting a numerical input. All commands are terminated with ‘\n’ or ‘;’
Command
*IDN?
Short
command
*IDN?
Type
Get
Purpose
Identify the unit. Reports the firmware version.
Input example
*IDN?
Output
example
[IDN](Hinds PEM 200 controller V01)
Command
:MODulator:DRiVe #
Short
command
:MOD:DRV #
Type
Get/Set
Purpose
Sets the voltage to the modulator on an arbitrary range of 0.0 (lowest
modulation amplitude) to 1.0 (highest modulation amplitude)
Input example
:MOD:DRV 0.1325
Output
example
[DRIVE](1.325000E-1)
Command
:MODulator:AMPlitude #
Short
command
:MOD:AMP #