PEM200 User Manual
36
Type
Get/Set
Purpose
Sets the voltage of the modulator in nanometers. For instance, if half wave
retardance at 633nm is desired, the argument should be 316.5
Input example
:MOD:AMP 316.5
Output
example
[AMP](3.2)
Command
:MODulator:AMPRange?
Short
command
:MOD:AMPR?
Type
Get
Purpose
Reports the modulation amplitude range in nanometers
Input example :MOD:AMPR?
Output
example
[AMPR](1.1,5.2)
Command
:MODulator:FREQuency?
Short
command
:MOD:FREQ?
Type
Get
Purpose
Reports the frequency of modulation in Hz
Input example
:MOD:FREQ?
Output
example
[FREQUENCY](5.014)