PEM200 User Manual
13
2
PEM-200 Controller
PEM-200 Controller Front Panel
The PEM-200 Controller front panel consists of six separate SMA connections:
HV:
Two SMAs enclosed in the high voltage (“HV”) box are for interface between optical
head and controller. These connections are necessary to drive the PEM optical head.
These are the high voltage SMA connections. Please ensure that the PEM200 controller
is off or disconnected from power when making any connection with these SMA ports.
Order (+/-) does not matter when connecting SMAs from HV output to optical head
inputs. See Operators Safety Summary for more detail.
Ref Out:
The “Ref Out” SMA will give a 3.3V TTL signal at the PEM’s selected
harmonic frequency for any triggering needs.
Trig Out:
Trig Out is a second reference SMA reserved for future expansion or custom
customer triggering needs upon request.
Signal In & Aux In:
Additional connections required for synchronous detection options.
Figure 2.1 PEM-200 Controller Front Panel