Box spring Hilo
TECHNICAL MANUAL
page 42 of 62
VII. MECHANICAL TECHNICAL DATA
Bed type
Hilo
Weight of an empty bed without options
125 kg
Degree of protection
IP X4
Safe working load
200 kg
Maximum patient weight
165 kg
Max. load on the ba knee rest
80 kg
Max. sound level during a manipulation
max 50dB(A)
* It should be taken into consideration that the save working load can only be realised, if the bed is
loaded as mentioned in the norm EN 60601-2-52 (i.e. backrest 45 %, fixed section 25 %, knee rest
and footrest 30 %). The bed may never be overloaded dynamically, in case all lifting power is
concentrated at one bed end. In this way, the bed is protected against structural deformation. This
lifting capacity applies to a bed load, with a user in it, and with accessories, mattress and options.
Furthermore the maximum load on the backrest by inclination is max 80 kg !
VIII. STORAGE OF THE BED
Before storing the bed, you have to make sure that:
▪
the bed and the mattress platform sections have been put in their lowest position.
▪
all electric functions have been switched off.
▪
all accessories have been removed.
▪
the optional battery has to be charged every week or disconnected from the control
box, in case the bed with a battery connected to the control box, is stored for more
than one week.
▪
the bed has been covered.
▪
the storage space is dry and dust-proof (relative humidity of 20 to 90% without
condensation).
▪
the temperature remains relatively constant and lies between -10 and +50°.
▪
the atmospheric pressure lies between 700 hPa and 1060 hPa
You have to check all functions before you want to use the bed again (see
checklist in appendix).
Summary of Contents for Hilo
Page 2: ...Box spring Hilo OPERATING INSTRUCTIONS page 2 of 62 ACTIVATION SUMMARY...
Page 3: ...Box spring Hilo OPERATING INSTRUCTIONS page 3 of 62 ASSEMBLY SUMMARY...
Page 34: ...Box spring Hilo TECHNICAL MANUAL page 34 of 62 TECHNICAL MANUAL...
Page 61: ...Box spring Hilo APPENDIX page 61 of 62 APPENDIX 3 TRAINING CERTIFICATE...