Preventive Maintenance
92
4.6.2.6 Internal Flow Sensor Calibration
Caution:
The three procedures “Proportional Valve Calibration”, “Internal Flow Sensor
Calibration” and “Characteristic of the Proportional Valve” must always be
performed together in the order as in the manual, never individually.
1
Select FLOW.INT to enter the Internal flow sensor screen sensor screen.
Fig. 42 Internal Flow Sensor Calibration Screen
2
Adjust minimum range calibration.
a
Select MIN to apply a zero flow.
b
Select MIN again to record the value.
3
Adjust maximum range calibration.
a
Select MAX to apply the maximum flow.
b
Select MAX again to record the value.
4
Select MIN again to close the valve.
5
Select SAVE to store the values into the memory.
6
Select QUIT to return to the main menu.