Preventive Maintenance
68
4.6 Navigation
1
While in the main menu screen, select individual keys to calibrate the O
2
sensor, Temperature
sensor, P1 pressure sensor, P2 pressure sensor, Internal flow sensor, External flow sensor or
proportional valve.
2
Select the test menu screen.
Fig. 24 Main menu screen