Repair Information
39
2.15.2.7.1 Status of the diaphragm/membrane valves:
The diaphragm valves are pneumatically operated by solenoid valves. The following list shows the
correspondence between diaphragm valve, solenoid valve and pressure port.
Valve
Bellows valve
Exp. valve
Decoupling valve
controlled by
MV 1
MV 2
MV 4
MV 3
connected to port #
5
1
3
6
wire #
5, 6
7, 8
1, 2
3, 4
In the following the status of the diaphragm/membrane valves are shown corresponding to the
selected ventilation mode of the ventilator.
2.15.2.7.2 Machine OFF or Ventilator In Standby
Valve
MV 1
MV 2
MV 3
MV 4
Activity
OFF
OFF
OFF
OFF
Pressure on port
5, No
1, No
6, No
no sup. of MV2
2.15.2.7.3 Manual / Spontaneous Mode, INSPIRATION and EXPIRATION
Valve
MV 1
MV 2
MV 3
MV 4
Activity
OFF
OFF
OFF
OFF
Pressure on port
5, No
1, No
6, No
no sup. of MV 2
2.15.2.7.4 CMV Child / Adult Mode, INSPIRATION
Valve
MV 1
MV 2
MV 3
MV 4
Activity
ON
ON
ON
ON
Pressure on port
5, Yes
1, Yes
6, Yes
supply of MV 2
2.15.2.7.5 CMV Child / Adult, EXPIRATION
Valve
MV 1
MV 2
MV 3
MV 4
Activity
OFF
OFF
OFF
OFF
Pressure on port
5, No
1, No
6, No
no sup. of MV 2
2.15.2.7.6 CMV Adult, EXPIRATION and PEEP
Valve
MV 1
MV2
MV3
MV4
Activity
OFF
ON
OFF
ON
Pressure on port
5, No
1, Yes
6, No
supply of MV2
Summary of Contents for Modular+
Page 1: ...HEYER Modular Anesthesia System Service Manual Rev 2 1 0 Software version Service software 1 8...
Page 12: ...Theory of Operation 12 This page intentionally left blank...
Page 26: ...Repair Information 26 2 14Ventilator pneumatic Fig 6 Ventilator pneumatic...
Page 90: ...Preventive Maintenance 90...
Page 98: ...Preventive Maintenance 98 Fig 45 How to measure the voltages U1 and U2...
Page 102: ...Preventive Maintenance 102 This page intentionally left blank...
Page 113: ...Order Information 113 This page intentionally left blank...