Preventive Maintenance
84
5
Adjust minimum range calibration
a
Apply ambient pressure to the system by removing the breathing circuit from the patient
module.
b
Turn the P2 OFFS potentiometer (P8, AVM 3-2) until the screen bar graph shows 20.
c
Select MIN on the touch screen.
6
Adjust maximum range calibration
a
Attach a Calibration pump (900-4726) to the Pressure sensor P2 on the Module 2, AVM 3-2
board.
b
Apply a pressure of 80 cmH
2
O to the Pressure sensor P2.
c
Turn the P2 GAIN potentiometer (P7, AVM 3-2) until the screen bar shows 80.
d
Select MAX on the touch screen.
e
Remove the calibration pump and reconnect the original tube to the pressure sensor P2
f
Check both valves alternately for correctness and readjust, if necessary.
7
Select SAVE twice on the touch screen to store minimum and maximum values in memory.
8
Select QUIT to return to the main menu
Caution:
Pressing QUIT at any time during the procedure will cancel the session’s
settings and reload the previously stored calibration coefficients.