Repair Information
18
2.9.1 Connectors on power supply board AVM 2-1
Connector
connected to
Pin. def.
Func. / Signal
Color / No.
X1
Module 1
X1-1
+5V
yellow
X1-2
GND
red
X1-3
GND
green
X1-4
+12V
blue
Connector
connected to
Pin. def.
Func. / Signal
Color / No.
X2
X2-1
+5V
X2-2
GND
X2-3
GND
X2-4
+12V
Connector
connected to
Pin. def.
Func. / Signal
Color / No.
X3
X2-1
-
X2-2
-
X2-3
GND
X2-4
+12V
Connector
connected to
Pin. def.
Func. / Signal
Color / No.
X4
X4-1
+5V
X4-2
GND
X4-3
GND
X4-4
+12V
Connector
connected to
Pin. def.
Func. / Signal
Color / No.
X5
Battery,
X5-1
BATT+
brown
Fuse for Battery
X5-2
BATT-
black
Main switch
X5-3
Cir.break. bat.
black 6
X5-4
Cir.break. bat.
black 5
X5-5
Main switch
black 4
X5-6
Main switch.
black 3
Connector
connected to
Pin. def.
Func. / Signal
Color wire isolat.
X6
Transformer
X6-1
sec.IIa 15VAC
blue
RT110/12
secondary
X6-2
sec IIa
green
X6-3
sec. IIb 15VAC
blue
X6-4
sec IIb
green
Connector
connected to
Pin. def.
Func. / Signal
Color wire isolat.
X7
Transformer
X7-1
prim.I, L120/230VAC black/
RT110/12
primary
X7-2
prim.I, N120/230VAC orange
X7-3
prim.II,L120/230VAC white
X7-4
prim. II, .
violet
Connector
connected to
Pin. def.
Func. / Signal
Col. wire isolat.
X11
Transformer
X11-1
Sec.Ia 22VAC
red
RT110/22
secondary
X11-2
Sec Ia
yellow
X11-3
Sec. Ib 22VAC
red
X11-4
Sec Ib
yellow
Summary of Contents for Modular+
Page 1: ...HEYER Modular Anesthesia System Service Manual Rev 2 1 0 Software version Service software 1 8...
Page 12: ...Theory of Operation 12 This page intentionally left blank...
Page 26: ...Repair Information 26 2 14Ventilator pneumatic Fig 6 Ventilator pneumatic...
Page 90: ...Preventive Maintenance 90...
Page 98: ...Preventive Maintenance 98 Fig 45 How to measure the voltages U1 and U2...
Page 102: ...Preventive Maintenance 102 This page intentionally left blank...