2.9.3 Charging / discharging control for the battery:
The charging / discharging control for the battery is also located on the power supply board AVM 2-1.
Status indicators as red, green and yellow LEDs are located on the board to show functions like check
or charging of the battery.
2.9.4 Status Indicators Of Battery Control:
LED
LED light
Battery status
LED yellow (V28)
continuous
charging
LED yellow (V28)
flashing
check
LED green (V37)
continuous
fully charged
LED red (V38)
flashing or continuous.
defect