Repair Information
40
2.15.2.7.7 Compliance Test Patient Module
The compliance test patient module is divided in 3 phases.
Phase 1. Flow into the re-breathing system and test of flow sensors.
Valve
MV 1
MV2
MV3
MV4
Activity
ON
OFF
ON
ON
Pressure on port
5, Yes
1, No
6, Yes
no sup. of MV2
Phase 2. Pressure increase up to test pressure 40 cmH
2
O, calculation of system compliance.
Valve
MV 1
MV2
MV3
MV4
Activity
ON
ON
ON
ON
Pressure on port
5, Yes
1, Yes
6, Yes
supply of MV2
Phase 3. Test of pressure loss.
Valve
MV 1
MV2
MV3
MV4
Activity
ON
ON
ON
ON
Pressure on port
5, Yes
1, Yes
6, Yes
supply of MV2
2.15.2.7.8 Leak Test Patient Module and Fresh Gas Module
Valve
MV 1
MV2
MV3
MV4
Activity
ON
OFF
OFF
OFF
Pressure on port
5, Yes
1, No
6, No
no sup. of MV2