![Hettich ROTINA 420 Repair Instructions Download Page 88](http://html1.mh-extra.com/html/hettich/rotina-420/rotina-420_repair-instructions_2127804088.webp)
87/97
10.13 Temperature sensor (B3) at the condenser (only in centrifuges with
cooling)
•
Remove the front panel and the support sheet for electronics (A1) as described in
chapters 10.1, 10.2, pg. 80.
•
Remove the upper part of the centrifuge housing as described in chapter "Removing
the upper part of the centrifuge housing".
•
Pull the plug S706 on the electronics (A1), see pg. 93, chapter 11.3.
•
Remove the temperature sensor (pg. 78, Fig. 5, item 37) at the condenser.
•
To mount the temperature sensor (B3), carry out these steps in opposite order.
The temperature sensor must be mounted on the condenser at exactly the same
position as earlier.
•
After the installation a compensation of the temperature sensor (B3) must be
performed, see pg. 55, chapter 6.14.4.
10.14 Fan (M3) at the condenser (only in centrifuges with cooling)
•
Remove the front panel and the support sheet for electronics (A1) as described in
chapters 10.1, 10.2, pg. 80.
•
Remove the upper part of the centrifuge housing as described in chapter "Removing
the upper part of the centrifuge housing".
•
Pull the plug S904 on the electronics (A1), see pg. 93, chapter 11.3.
•
Unscrew the two fastening screws of the fan.
•
Remove the fan (pg. 78, Fig. 5, item 38).
•
To mount the fan (M3), carry out these steps in opposite order.
10.15 Fan (M5) at the motor (only in centrifuges with cooling)
•
Remove the front panel and the support sheet for electronics (A1) as described in
chapters 10.1, 10.2, pg. 80.
•
Remove the motor as described in chapter "Removing Motor (M1) / Rubber-metal
bearings / Speed sensor (B2.1, at the bottom of the motor) / Imbalance sensor
(B2.2)".
•
Pull the plug S1001 on the electronics (A1), see pg. 93, chapter 11.3.
•
Unscrew the two fastening screws of the fan.
The two fastening screws are only accessible through the centrifuge
chamber.
•
Remove the fan (pg. 78, Fig. 5, item 39) through the centrifuge chamber.
•
To mount the fan (M5), carry out these steps in opposite order.
Summary of Contents for ROTINA 420
Page 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Page 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Page 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Page 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Page 96: ...95 97 ...
Page 97: ...96 97 ...