71/97
Action
Display / Comment
4. Press the
PROG
key
so often until the
following will be
displayed:
→
EE Init Global Data no
•
Turn the control
knob one step to
the right.
→
EE Init Global Data yes
no = Do not initialise the EEPROM.
yes = Initialise the EEPROM.
5. Press the
START
key.
→
Are you sure ? no
Repeated confirmation.
•
Turn the control
knob one step to
the right.
→
Are you sure ? yes
no = Do not initialise the EEPROM.
yes = Initialise the EEPROM.
6. Press the
START
key.
→
-
>
Eeprom
7. Press the
STOP / OPEN
key.
→
∗∗∗
Select Menu
∗∗∗
8. Press the
STOP / OPEN
key.
→
→
→
→
→
EE INIT : Global Data …
The EEPROM is initialised and afterwards the "Select
Menu" will be terminated.
e.g.
ROTINA 420R
Centrifuge model
e.g.
V 01.01
Program version
e.g.
Rotor 4 Nmax=4500 R=184
Rotor information
e.g.
6 20 9
9 4000 5:20
Display of the centrifugation data.
6.21 Automatic
repetition
of the centrifugation run
For test purposes the automatic repetition of the centrifugation run can be selected.
The number of centrifugation runs is counted automatically.
The break time between the centrifugation runs can be set.
This automatic repetition remains activated until the test run is terminated or the
centrifuge will be switched off.
The setting is only possible if the rotor is at standstill.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
key.
Summary of Contents for ROTINA 420
Page 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Page 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Page 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Page 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Page 96: ...95 97 ...
Page 97: ...96 97 ...