![Hettich ROTINA 420 Repair Instructions Download Page 54](http://html1.mh-extra.com/html/hettich/rotina-420/rotina-420_repair-instructions_2127804054.webp)
54/97
6.14.3
Deactivate the temperature sensor (B2) at the top of the centrifuge
chamber
The ROTINA 420 R / 420 RH has no temperature sensor (B2) at the top of the
centrifuge chamber and therefore the temperature sensor (B2) must be absolutely
deactivated in the program. The temperature sensor (B2) is ex factory deactivated.
Only with centrifuges that have at the top of the centrifuge chamber the temperature
sensor (B2) this sensor must be activated in the program.
The deactivation of the temperature sensor (B2) is only possible if the rotor is
at standstill.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
key. In this case the adjustments are not saved.
Action
Display / Comment
1. Keep the
START
key
and the
STOP / OPEN
key pressed and
switch on the mains
switch at the same
time.
→
∗∗∗
Select Menu
∗∗∗
2. Press the
PROG
key
so often until the
following will be
displayed:
→
-
>
Setup
3. Press the
START
key.
→
OEM Version = off/on
4. Press the
PROG
key
→
T2 Sensor = --- / ok.
--- = Temperature sensor deactivated.
ok. = Temperature sensor activated.
•
Turn the control
knob .
→
T2 Sensor = ---
Set "---" because the ROTINA 420 R has no temperature
sensor B2 (= T2).
5. Press the
START
key.
→
→
Store Setup ...
The setting will be stored.
-
>
Setup
6. To exit the menu
press the
STOP / OPEN
key twice.
→
→
∗∗∗
Select Menu
∗∗∗
e.g.
6 20 9
9 4000 5:20
Display of the centrifugation data.
Summary of Contents for ROTINA 420
Page 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Page 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Page 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Page 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Page 96: ...95 97 ...
Page 97: ...96 97 ...