21/97
ROTORCODE 10.3
Error during reading the rotor code.
The rotor slows down braked until it stops.
No further user operation possible except opening the lid.
•
Reset error code:
Perform a MAINS RESET.
•
Magnetic code on the rotor is defective, see pg. 91, chapter 11.1.
•
Electronics (A1) is defective.
MAINS INTERRUPT
Interruption of mains supply during centrifugation.
During the interruption of the mains supply the rotor slows down
without braking until it stops.
After the interruption of the mains supply the rotor slows down
braked until it stops.
•
Reset error code:
Open the lid and press the
START / IMPULS
key.
The error code
cannot
be reset by a MAINS RESET.
•
Power failure.
•
Overtemperature switch (F3) on the brake resistor (R1) has
opened or is defective.
•
Loose contact in the electrical wiring.
•
Electronics (A1) is defective.
VERSION ERROR 12
Incorrect machine type and/or cooling version identified.
No further user operation possible except opening the lid.
•
Perform an initialisation of the EEPROM, see chapter "6.20".
Then check the machine type and the cooling version (dispaly
after MAINS RESET).
•
The installed electronics (A1) does not fit to the centrifuge model.
•
Electronics (A1) is defective.
Summary of Contents for ROTINA 420
Page 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Page 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Page 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Page 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Page 96: ...95 97 ...
Page 97: ...96 97 ...