70/97
Action
Display / Comment
•
Turn the control
knob to the right.
→
e.g.
Adr: 0010 = 000003E3 - 00000215
Content of the memory cells 10, 12, 14, 16, 18, 1A, 1C, 1E.
Example: Content of the address 16 = E3
By turning the control knob the contents of the memory
cells 0000 up to 7FF0 can be displayed.
If the control knob is turned to the right and the
PROG
key is pressed at the same time then the
address will change in steps of 1000
hex
.
5. To exit the menu
press the
STOP / OPEN
key four times.
→
→
→
→
Eeprom Read
-
>
Eeprom
∗∗∗
Select Menu
∗∗∗
e.g.
6 20 9
9 4000 5:20
Display of the centrifugation data.
6.20 Initialise
the
EEPROM
If the EEPROM is initialized, the centrifuge model, the setting of the imbalance
switch-off and the offset values of the temperature sensors are deleted. The
centrifuge model, the offset-values and the imbalance switch-off must be set
again after the initialization. Recommendation: Note the adjusted imbalance
value and the offset-values of the temperature sensors before initializing and
enter afterwards again.
The initialisation of the EEPROM is only possible if the rotor is at standstill.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
key.
Action
Display / Comment
1. Keep the
START
key
and the
STOP / OPEN
key pressed and
switch on the mains
switch at the same
time.
→
∗∗∗
Select Menu
∗∗∗
2. Press the
PROG
key
so often until the
following will be
displayed:
→
-
>
Eeprom
3. Press the
START
key.
→
Eeprom Read
Summary of Contents for ROTINA 420
Page 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Page 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Page 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Page 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...