34/97
N > ROTOR MAX
Speed in the selected program greater than the maximum speed of
the rotor.
No start possible.
•
Check and reduce the set speed.
FC INIT ERROR
Initialisation
of the frequency converter is faulty.
No further user operation possible except opening the lid.
•
Reset error code:
Perform a MAINS RESET.
•
Electronics (A1) is defective.
FATAL EEPROM ERROR 1 – 5
EEPROM
error.
No further user operation possible except opening the lid.
•
Reset error code:
Perform a MAINS RESET.
•
Initialise the EEPROM, see pg. 70, chapter 6.20.
•
Electronics (A1) is defective.