EN
54/83
9
Unpacking the centrifuge
Lift the carton upward and remove the padding.
Observe the weight of the centrifuge, refer to chapter "Technical specifications".
With a suitable number of helpers, hold the centrifuge on both sides and lift down from the pallet.
10
Removing the transport securing device
It is imperative that the transport securing device be removed.
Keep the transport securing device in a safe place since it must be installed again before transporting the
device.
The device may only be transported with the transport securing device installed.
To protect the device from damage during transport, the motor is fixed in place.
This transport securing device must be removed when the device is put into operation.
ROTINA 380 Robotic
(a)
(c)
(d)
(d)
(a)
(b)
(c)
(b)
Open the lid.
Screw out the four screws (a) and remove the motor cover (b).
Remove the three screws (c) and spacer sleeves (d).
Turn the motor cover (b) over and fasten to the motor with the four screws (a).
The transport securing device is installed in the opposite order.