HE 5750 Operating Instructions
#371627 | Version 1.1
75
12.4 Diagnosis HE 5750 at PLC
Position
Bit
Function
Data type
Bytes
0
System status used by Profibus SPC3 chip
EXT_USER_DIAG
unsigned
char
6
6
1. Header
Digital
1
7/6
00 = device specific diagnosis
5-0
Length = 11 Byte
7
Status
Digital
1
0
invalid parameter data
unsigned
char
1
Invalid configuration
2
Invalid command
3
Invalid test
8
I/O-extension status (HE 5910)
Digital
1
0
I/O-Unit 1 not connected
unsigned
char
1
I/O-Unit 1 error
2
I/O-Unit 2 not connected
3
I/O-Unit 2 error
4
I/O-Unit 3 not connected
5
I/O-Unit 3 error
6
I/O-Unit 4 not connected
7
I/O-Unit 4 error
9
I/O-extension status (HE 5910)
Digital
1
0
I/O-Unit 5 not connected
unsigned
char
1
I/O-Unit 5 error
2
I/O-Unit 6 not connected
3
I/O-Unit 6 error
4
I/O-Unit 7 not connected
5
I/O-Unit 7 error
6
I/O-Unit 8 not connected
7
I/O-Unit 8 error
10
I/O-extension status (HE 5910)
Digital
1
0
I/O-Unit 9 not connected
unsigned
char
1
I/O-Unit 9 error
2
I/O-Unit 10 not connected
3
I/O-Unit 10 error
11
I/O-extension status (HE 5910)
Digital
1
12
reserved
Chamber related
5
17
2. Header
Digital
1
7/6
00 = device specific diagnosis
5-0
Length = 33 Byte
Chamber 1
2
nd
1
Slave Status
Digital
1
0
Slave 1 not connected
unsigned
char
1
Slave 1 Error
2
Slave 2 not connected
3
Slave 2 Error
2
nd
2
Pressure alarms
Digital
1
0
Manifold sensor 1 wire break
unsigned
char