![Heraeus CRYOFUGE 6000i Operating Instructions Manual Download Page 33](http://html1.mh-extra.com/html/heraeus/cryofuge-6000i/cryofuge-6000i_operating-instructions-manual_2125625033.webp)
"SEPACONTROL" Programming
29
Program memory operation
Creating and saving standard programs (position no. "01".... "32")
Key switch position
Changing programs is only possible in key switch position A.
CAUTION: The centrifugation process must not have been started!
Pre-setting a program number
With the directional program selection keys, the desired number from 01 to 32 or profile program (F1,
F2,…) can be selected in single steps.
Activating programming
The pre-set program number can be displayed flashing by pressing the "store mode"
key. This visually indicates that an adjustment of the program set values is possible.
Entering set values
The set values are to be entered as described under “RAM memory operation“.
Ending programming
The entered set value record can be stored at the chosen program memory position by
pressing the "execute" key thereby ending the flashing phase.
Creating and saving a profile program (position "F", max. 10 levels)
The profile program enables the user to establish an individual program from acceleration and deceleration
profiles as well as speed levels with a maximum of 10 different segments.
This additional program is available in the user memory in position 33 under display "F".
Key switch position
Changing profile programs is only possible in key switch position A.
CAUTION: The centrifugation process must not be started!
Pre-setting a program number
Call up the memory segment "F1" using the directional program selection keys.
Activating programming
The memory segment "F1" is displayed flashing by pressing the "store mode" key.
Entering set values
In the segment "F1", all parameters of the complete centrifuging process must be set (bucket order number
and set temperature cannot be changed in the following levels). With the "upward arrow" key, the values set
are saved and the transition is made to "F2". For all the following segments ("F2" - "F0"), only the accelera-
tion or deceleration curves, speed and time values can be set. For the last segment, the deceleration curve
number (0-9) must be selected and the speed set to zero.
Attention!
After completion of a programmed time segment, the appliance will switch to the next profile segment inde-
pendent of the speed value set. If all 10 available program segments are occupied, the appliance will decel-
erate using deceleration curve 9 after completion of segment 10.
Ending programming
The program set is stored in the profile program memory position by pressing the
"execute" key thereby ending the flashing phase.
store
mode
execute
program
execute
store
mode
execute
program
execute
Summary of Contents for CRYOFUGE 6000i
Page 1: ...CRYOFUGE 6000i CRYOFUGE 8500i Operating Instructions ...
Page 4: ......
Page 8: ...Safety Measures and Hazards 4 ...
Page 12: ...Description of Appliance 8 ...
Page 14: ...Instructions for Rapid Start up 10 ...
Page 22: ...Transport and Installation 18 ...
Page 42: ...Instructions for Temperature Control 38 ...
Page 46: ...Maintenance and Service 42 ...