"SEPACONTROL" Programming
27
Sequential control
The control of the centrifuge is possible in any position of the key switch.
The release of the control key functions is indicated by the current standby LED on the right next to the key.
Opening the lid
Only when the centrifuge is switched on and the rotor is not turning any more and no
error message is shown, can the lid be opened using the lid symbol key (yellow LED
is alight). When the lid is open, the text "OPEN" is displayed in the multi-functional
field. The start function is now blocked.
Starting
When the start release takes place (green LED is alight), centrifugation can be initi-
ated using the "start/stop" key. After the start, the lid symbol key is blocked (yellow
LED is off).
Stopping
After each start the stop release (red LED is alight) is immediately activated. By
pressing the "start/stop" key, centrifugation is terminated.
After each stop, the green LED is alight again. This means that the interrupted cen-
trifuging process can be continued at any time (it is not necessary to wait until the
rotor is at a standstill, in case it was unintentionally stopped).
Quick stop
Each centrifugation can also be stopped using the quick stop symbol key, when the
red LED is alight. The "quick stop" function has been developed as the quickest form
of deceleration.
Attention!
Pressing the quick stop key does not overwrite the programmed deceleration curve!
start
stop
start
stop
Summary of Contents for CRYOFUGE 6000i
Page 1: ...CRYOFUGE 6000i CRYOFUGE 8500i Operating Instructions ...
Page 4: ......
Page 8: ...Safety Measures and Hazards 4 ...
Page 12: ...Description of Appliance 8 ...
Page 14: ...Instructions for Rapid Start up 10 ...
Page 22: ...Transport and Installation 18 ...
Page 42: ...Instructions for Temperature Control 38 ...
Page 46: ...Maintenance and Service 42 ...