Technical Specifications
Dimensions: 85 × 160 × 43 mm (3.35 × 6.3 × 1.69 inches)
Weight: 350 g (0.77 lbs) with batteries installed
4 batteries type AA or R6, 1.5 V
Power adapter: CE marked
Only use adapters recommended by HemoCue, as listed
under “Recommended Adapters”.
Pollution degree: 2
Atmospheric pressure: 700 hPa to 1060 hPa
Overvoltage category: II
Equipment not suitable for use in the presence of flammable
mixtures. The instrument is tested according to IEC 61010-1,
Second edition: 2001 and EN 61010-1: 2001, IEC/EN 61010-
2-101: 2002, IEC 61326-1: 2005 and EN 61326-1: 2006, IEC
61326-2-6, First edition: 2005 and EN 61326-2-6: 2006, IEC
60601-1-2, Third edition: 2007 and EN 60601-1-2: 2007 and
complies with the IVD Medical Device Directive 98/79/EC.
The instrument is made for continous mode.
Recommended Adapters
Marked HCA01
Input: 100–240 V ~/50–60 Hz. <500 mA
Essential Performance
The essential performance is quantitative determination of
glucose in capillary, venous or arterial whole blood.
Warranty
The analyzer carries a 24-month warranty from the day of
receipt. After the expiration date of the warranty, main-
tenance and repairs are offered at a fixed price. For tech-
nical difficulties or repair contact your local distributor or
HemoCue AB. Any other use of the system than recommended
by the manufacturer will void the warranty.
Service and Disposal
The analyzer should be cleaned as recommended under
Maintenance prior to service or disposal. Consult your local
environmental authorities for proper disposal.
Spare parts – Accessories
The following spare parts and accessories are available:
• AC Power adapter
• Battery Lid
• Cuvette holder
• HemoCue Cleaners
• Hard Carrying Case
• HemoCue Lancets
Summary of Contents for Glucose 201 RT
Page 1: ...Gebruiksaanwijzing Manuel d utilisation Bedienungsanleitung Operating Manual ...
Page 2: ...2 ...
Page 6: ...2 4 1 3 6 1 3 5 6 Components Bestandteile Composants Onderdelen ...
Page 8: ...1a 3 1c 4 1b 4 4 2 8 Start up Inbetriebnahme Démarrage Opstarten ...
Page 14: ...5 6 7 8 14 ...
Page 16: ...9 10a 12 11 10b 12 12 16 ...
Page 18: ...18 15 16 15 14 mg dl mmol l 13 40 s ...
Page 22: ...22 4 1 2 5 6 3 Set up QC test Einstellung QC Test Configuration Test QC Set up QC test ...
Page 98: ...98 ...
Page 99: ...99 ...
Page 100: ...100 ...
Page 101: ...101 ...
Page 102: ...102 ...
Page 103: ...103 ...