53
Specific performance characteristics
Repeatability and within-laboratory precision
Repeatability and within-laboratory precision was determined
according to the CLSI Document EP5. The results given below
in “Repeatability” and “Within-laboratory precision” come
from 2 batches of HemoCue Glucose 201 RT Microcuvettes,
10 HemoCue Glucose 201 RT Analyzers and 6 operators.
Commercially available controls at 3 different levels were
used. The glucose concentration was measured in duplicate
twice a day, morning and afternoon, during 20 working days.
Repeatability
Within-laboratory
precision
Control
level
N
Mean
mg/dL
(mmol/L)
SD
pooled
mg/dL
(mmol/L)
CV
%
SD
pooled
mg/dL
(mmol/L)
CV
%
1
800
45.9 (2.5)
1.15 (0.064)
2.5
1.28 (0.071)
2.8
2
800
101.8 (5.7)
1.36 (0.075)
1.3
1.62 (0.090)
1.6
4
800
305.7 (17.0)
3.26 (0.180)
1.1
3.72 (0.210)
1.2
Validation Studies
Results of the validation studies, based on the intended use
of the system, are shown in Figure 1. Single replicate on the
HemoCue Glucose 201 RT system and the mean of duplicate
on the comparative method are presented in the plots. All
results are in whole blood.
Figure 1
Users Performance Evaluation
Point-of-Care
Capillary samples
Coefficients:
y
0
= 2.152
a = 1.002
r
2
= 0.995
n = 102
G
lu
co
se 2
01 R
T, 1
st r
ep
l (
m
g
/d
L)
YSI 2300 STAT Plus, mean values (mg/dL)
Summary of Contents for Glucose 201 RT
Page 1: ...Gebruiksaanwijzing Manuel d utilisation Bedienungsanleitung Operating Manual ...
Page 2: ...2 ...
Page 6: ...2 4 1 3 6 1 3 5 6 Components Bestandteile Composants Onderdelen ...
Page 8: ...1a 3 1c 4 1b 4 4 2 8 Start up Inbetriebnahme Démarrage Opstarten ...
Page 14: ...5 6 7 8 14 ...
Page 16: ...9 10a 12 11 10b 12 12 16 ...
Page 18: ...18 15 16 15 14 mg dl mmol l 13 40 s ...
Page 22: ...22 4 1 2 5 6 3 Set up QC test Einstellung QC Test Configuration Test QC Set up QC test ...
Page 98: ...98 ...
Page 99: ...99 ...
Page 100: ...100 ...
Page 101: ...101 ...
Page 102: ...102 ...
Page 103: ...103 ...