35
Symptom
Explanation
Action
E35
1. The current measurement has been incorrectly
stored and will remain as error code E35 until
the entire memory has been deleted
1a. No action is normally required.
1b. If storage problems continue, the analyzer needs service.
Contact your local distributor.
E70–E71
Erroneous kinetics.
1. The cuvette is faulty.
2. The sample might be grossly lipemic.
1–2. Check that the system is used according to the instru ction
for use.
Fill a NEW cuvette and place it immediately in the analyzer
If the error code appears again, use a suitable laboratory refe-
rence method to analyze the specimen.
ECU
1. The cuvette is empty or not filled with enough
sample. Please note, the ECU function captures
primarily empty cuvettes, not under-filled
cuvettes.
2. No chemical reaction is identified in the blood
filled cuvette.
1–2. Fill a NEW cuvette and place it immediately in the analyzer.
If the ECU function appears again, contact your local distributor
or HemoCue AB. Use another HemoCue analyzer or a suitable
laboratory reference method to analyze the specimen.
HHH
1. Measured value exceeds 27.8 mmol/L (500 mg/
dL) for a whole blood system and 31 mmol/L
(560 mg/dL) for a plasma equivalent system.
1. The result is above the measuring range.
Summary of Contents for Glucose 201 RT
Page 1: ...Gebruiksaanwijzing Manuel d utilisation Bedienungsanleitung Operating Manual ...
Page 2: ...2 ...
Page 6: ...2 4 1 3 6 1 3 5 6 Components Bestandteile Composants Onderdelen ...
Page 8: ...1a 3 1c 4 1b 4 4 2 8 Start up Inbetriebnahme Démarrage Opstarten ...
Page 14: ...5 6 7 8 14 ...
Page 16: ...9 10a 12 11 10b 12 12 16 ...
Page 18: ...18 15 16 15 14 mg dl mmol l 13 40 s ...
Page 22: ...22 4 1 2 5 6 3 Set up QC test Einstellung QC Test Configuration Test QC Set up QC test ...
Page 98: ...98 ...
Page 99: ...99 ...
Page 100: ...100 ...
Page 101: ...101 ...
Page 102: ...102 ...
Page 103: ...103 ...