Evolution™ Electric Bed LI156Ex
QD1532A(2)
51
USER MANUAL
S
A
F
E
T
Y
T
IP
S
AND
P
R
E
CAUT
IO
NS
7.2.2 PREVENTIVE MAINTENANCE
An annual maintenance visit is recommended to keep the bed fully
operational and to ensure optimal use.
The following points should be given particular attention:
- Brake locking system
- Siderail locking mechanisms
- Drive systems (especially jacks), movement control systems (Head
Section, Leg section and Trendelenburg / Reverse Trendelenburg) and
their ancillary parts
- Bed movement and ancillary part bearings
- Control operation (in particular return to neutral)
- The state of cables and electrical components
- Correct earthing of metal parts and equipotential connectors
- Waterproofing of electrical connections
- Cabling
- Correct condition of the frame and welded assemblies (corrosion and
shocks).
The whole electrical system should be inspected by an approved after-sales
service on average once every three years in order to ensure continued
operation in optimal conditions.
The frequency of the maintenance visits should be determined according
to the condition of the bed and its specific usage, as some components may
need changing after a given period of use (see Maintenance Manual).