Evolution™ Electric Bed LI156Ex
26
QD1532A(2)
USER MANUAL
IN
S
T
RUCT
IO
NS
F
O
R
U
S
E
4.12 RESTRAINING STRAP POSITIONS
When
required,
patient-restraining
straps must be tied to the proper
positions provided.
Two positions are located on each
bed section (Head Section, Knee
Section and Leg Section) on either
side of the bed.
Thread
the
straps
through
the
positions.
Do not fix the restraining straps to any position (particularly siderails)
other than those provided for this purpose.
When the patient is immobilised by the straps, the electrical functions must
be locked out and the bed must be fitted with either accessory AD113A
(Siderail protective cover) or a system of ankle restraints must be used.
4.13 FOOT END BED EXTENSION (optional)
To use the extension:
- Lift
the
handle
located
under
the
footboard
- Pull out the extension.
Three positions are available:
- 6 cm - 12 cm - 18 cm -
The extension locks into position with a
safety catch.
When extension is no longer in use, use the
same method to store the extension away.
A specially fitted panel and cushion are supplied as accessories with the
extension (see section 5.4).
Restraining strap positions
Knee Section
Restraining strap positions
Head section and Leg
section