Evolution™ Electric Bed LI156Ex
QD1532A(2)
9
USER MANUAL
IN
STA
LLIN
G
TH
E
B
ED
3. INSTALLING THE BED
3.1
FIRST STEPS
Before using the bed, it is essential to have a thorough understanding of this
manual. It contains instructions for general use and maintenance and gives
you the assurance of improved safety.
Nursing staff must be informed of the risks encountered in the use of electric
beds.
The many sources and types of accessories, hardware or medical devices
that may be used jointly with this bed do not enable Hill-Rom to guarantee
both the safety and conformity of all the combinations thus created. The
operator who creates these device combinations must therefore ensure that
security and conformity requirements are met.
Before installing the bed for the first time or after bringing the bed and its
accessories out of storage:
-
Ensure that the bed and its various parts are at room temperature
-
Connect the bed to the mains power supply (see sect. 3.2 below)
-
Ensure that all the moving parts are in good working order
-
Ensure that the bed has been cleaned and disinfected (see chapter 6).
3.2
CONNECTION TO THE MAINS POWER SUPPLY
The mains power supply for the bed must comply with French standards NF
C 15-100 et NF C 15-211 (230V power point + earth) (France) and CEI 364
for other destinations.
Check that the power supply of the bed (identification label – sect. 2)
corresponds to the power supply voltage of the hospital.
The power supply should be equipped with a maximum 30 mA earth leakage
circuit breaker, in compliance with CEI 364-5-53.