8.7 Bath Electrode
51
The body of the electrode is a 1 ml plastic syringe body that has been heated and pulled to form a small, bent
tip. The electrode proper is a chlorinated Ag wire that is first inserted with the plunger into the fluid-filled body;
then hot agar is sucked into the tip by withdrawing the plunger partially. The filling solution can be a typical bath
solution or something similar, such as 150 mM NaCl. More concentrated salt solutions are not necessary, and they
can leak out, changing the composition of the bath solution.