5.1 EPC 10 USB Window
29
current potential has been set correctly). The ”V-membrane” label is converted to ”I-membrane” in current clamp
mode.
I-mon:
Displays the actual measured pipette current.
Figure 5.4: Input monitor
V-mon:
Displays the actual measured pipette voltage after correcting for liquid junction potentials and offsets
(provided the zero-current potential has been set correctly). This may differ (temporarily) from the holding voltage
(e.g., during long stimulation pulses) as it indicates the average sum of
V-membrane
and the scaled stimulus voltage.
R-memb:
The seal resistance (
R-membrane
) is determined from the current sampled during the baseline and the
second half of the
Test Pulse
.
R-membrane
can be encoded into a tone using the
Sound
feature (see below).
Input ADC:
The
Oscilloscope
can display the following signals (
EPC 10 USB
Single):
Figure 5.5: Input ADC
Imon1: Current monitor 1 (after
Filter 1
)
Imon2: Current monitor 2 (after
Filter 2
)
Vmon: Voltage monitor output
AD 0. . . 4: Any of the other AD channels.
Recording Mode:
Sets the Recording Mode. For details on the various recording modes see chapter 6 on page 39.
Figure 5.6: Recording Mode
In Out: Sets the Inside Out mode.
On Cell: Sets the On Cell mode.
Out Out: Sets the Outside Out mode.
Whole Cell: Sets the Whole Cell mode.
C-Clamp: Sets the Current Clamp mode.
Test Pulse:
Test Pulses
are added to the holding potential and applied to the pipette; the current responses
are sampled and displayed. Two built-in
Test Pulse
types are available: single or double pulse. Additionally any
user-defined pulse pattern can be used as a
Test Pulse
.
Test Pulses
are applied at maximal rates depending on the
duration specified.
Amplitude / Length:
Duration and amplitude of built-in
Test Pulses
can be specified in the dialog. The
minimum pulse duration is 0.1 ms with 1000 points sampled per pulse.
Figure 5.7: Test Pulse parameters
http://www.heka.com
Summary of Contents for EPC 10 USB
Page 1: ...Hardware Manual Version 2 8 EPC 10 USB Computer controlled Patch Clamp Amplifier...
Page 6: ......
Page 10: ...4 Introduction http www heka com...
Page 16: ...10 Description of the Hardware http www heka com...
Page 22: ...16 Installation http www heka com...
Page 32: ...26 Verifying and Testing the EPC 10 USB http www heka com...
Page 44: ...38 The control software http www heka com...
Page 48: ...42 Operating Modes http www heka com...
Page 54: ...48 Compensation Procedures http www heka com...
Page 58: ...52 Patch Clamp Setup http www heka com...
Page 64: ...58 Using the Patch Clamp http www heka com...
Page 74: ...68 Appendix II Probe Adapters http www heka com...
Page 76: ...70 Appendix III S Probe http www heka com...
Page 81: ......