
5.1 EPC 10 USB Window
31
In the upper box, the total
C-fast
value is displayed.
τ
-fast
determines the time constant of
C-fast
(up to 8
µ
s).
The value of
τ
-fast
may be adjusted by dragging the mouse, or typing, or automatically by selecting the
Auto
function.
<>
: This will select both amplitude and time constant of
C-fast
. Values for each may be changed manually
depending on whether the mouse is moved horizontally or vertically.
Figure 5.9: C-Fast compensation
Auto C-fast:
Selection of this button performs an automatic compensation of
C-fast
and
τ
-fast
. The procedure
uses a routine that applies a number of small pulses (5 mV), averages the resulting currents and fits an exponential to
deduce the capacitance compensation values required to cancel the current. During this procedure some parameters
are changed temporarily:
Rs Comp
is turned off, the stimulation is interrupted and the
Gain
is internally set to
an appropriate value (0.2, 2 or 50 mV/pA, depending on the effective
Gain
range). The values are updated in the
respective displays. If the automatic compensation routine fails to converge to ideal adjustment, the small box in
the
Auto
button (E) will be darkened to indicate the error. Auto compensation works best with Sylgard
-coated
pipettes in the cell-attached configuration with
C-slow
turned off. The software allows for two ways of making the
compensation. The fastest (default setting) is to use a lookup table for setting
C-fast
(
EPC 10
menu
→
C-fast
Mode
→
Use Lookup Table
). Here, the lookup table created during the last calibration will be used. The slower
method is
Computed
(same menu) where an iterative compensation is performed.
C-slow:
This is used to cancel slow capacitive currents that charge the cell membrane in the whole-cell con-
figuration. The 30, 100 and 1000 pF ranges actually allow capacitance values to be compensated in the ranges
of 0.12-30 pF, 0.4-100 pF and 4-1000 pF, respectively. The adjustment range is also limited by the program in
order to make the time constant
R-series
*
C-slow
greater than 5
µs
to prevent oscillations. The
C-slow
value
may be adjusted manually by dragging the mouse, or typing, or automatically by selecting the
Auto
function.
Auto-compensation will adjust
C-slow
as well as
R-series
.
C-slow Range:
Selects the range for slow capacitance compensation:
Off: Turns cancellation off.
30 pF: Small cells.
100 pF: Small and medium-sized cells.
1000 pF: Large cells (low and intermediate gain range only).
Slow capacitance cancellation ranges (30/100/1000 pF) can be set to any desired value. However, in the high
Gain
range (50 GΩ resistor) the 1000 pF range will not operate. If the
Gain
is set to value higher than 20 mV/pA,
while the 1000 pF range is selected, the capacitance cancellation range will automatically be reduced to the highest
possible setting (100 pF). Similarly, since the current clamp mode is not possible in the high
Gain
range (50-
2000 mV/pA), the
Gain
will be reduced appropriately when selecting
C-Clamp
mode from high
Gain
range.
The program will prevent oscillations by limiting compensation according to the selected range and the effective
capacitance values (
R-series
*
C-slow
>
5
µs
). Selecting the 1000 pF range will limit the
Gain
adjustment to a
maximum of 20 mV/pA.
Note:
If you are using the
Cap. Track
feature (
Windows -> Configuration -> Display -> Show
Cap. Track
) to measure small capacitance changes, it is advantageous to use a higher
C-slow
range
when possible. Counter-intuitive as it may seem, the digital control of the
C-slow
circuitry is more
precise in higher ranges, resulting in higher resolution, consistent with the
Gain
range and capacitance.
Figure 5.10: C-slow Compensation
http://www.heka.com
Summary of Contents for EPC 10 USB
Page 1: ...Hardware Manual Version 2 8 EPC 10 USB Computer controlled Patch Clamp Amplifier...
Page 6: ......
Page 10: ...4 Introduction http www heka com...
Page 16: ...10 Description of the Hardware http www heka com...
Page 22: ...16 Installation http www heka com...
Page 32: ...26 Verifying and Testing the EPC 10 USB http www heka com...
Page 44: ...38 The control software http www heka com...
Page 48: ...42 Operating Modes http www heka com...
Page 54: ...48 Compensation Procedures http www heka com...
Page 58: ...52 Patch Clamp Setup http www heka com...
Page 64: ...58 Using the Patch Clamp http www heka com...
Page 74: ...68 Appendix II Probe Adapters http www heka com...
Page 76: ...70 Appendix III S Probe http www heka com...
Page 81: ......